Understanding ICDs
Implantable Cardioverter-Defibrillators (ICDs) are devices that look similar to standard pacemakers; however, they are also used to treat dangerous, rapid heart rhythms (tachyarrhythmias). Certain abnormal heartbeats starting in the lower chambers of the heart (ventricular arrhythmias) can become life-threatening. This happens because they cause the heart to beat excessively fast, preventing it from filling properly with blood during each contraction. An ICD is capable of immediately recognizing these irregular rhythms and promptly intervening to treat them.
The key difference between an ICD and a standard pacemaker is that the ICD includes a specialized wire placed in the right lower chamber of the heart—the defibrillation lead. In addition to providing regular pacing support, this lead features one or two electrical coils. Working together with the ICD device, these coils can deliver a high-energy electrical shock (up to 46 Joules) to stop a dangerous, rapid heart rhythm, such as ventricular fibrillation. This lifesaving process is known as defibrillation.
How an ICD Works
ICDs can function as standard pacemakers if the heart needs help beating faster (pacing support). If no pacing is needed, the device simply monitors the patient’s natural heart rhythm. Its primary function, however, is to instantly manage rapid heartbeats within seconds, before they pose an imminent threat to the patient. The device can intervene in two distinct ways.
The first method is called Antitachycardia Pacing (ATP), which is used for stable, rapid heartbeats. It involves delivering rapid electrical pulses (not shocks) at a rate faster than the abnormal rhythm, aiming to override and stop it. ATP intervention is completely painless and typically goes unnoticed by the patient.
If ATP fails to stop the rapid rhythm, the device uses its second method: delivering an electrical shock. By effectively resetting the chaotic electrical activity of the heart, this shock allows the normal heart rhythm to return. The patient feels the shock as a sudden, intense internal jolt. Although a shock can be a lifesaving intervention, it is painful and distressing; therefore, doctors must take all necessary measures to prevent it from happening.
Who Needs an ICD?
ICDs are recommended for patients who have either survived a dangerous, life-threatening irregular heartbeat episode or are at high risk of experiencing such an event in the future, such as individuals with severe heart failure. For example, patients who have suffered a past heart attack (myocardial infarction) and have a low ejection fraction—meaning the left lower chamber of the heart pumps out less blood than normal—face an increased risk of severe rhythm problems.
In a landmark clinical trial known as MADIT II, patients with a history of a heart attack and a low ejection fraction (≤ 30%) were randomly split into two groups: one group received an ICD, while the other followed standard conservative medical treatment. The study clearly demonstrated that patients who received an ICD had a higher life expectancy, precisely because the device saved them from fatal irregular heartbeats.
The Implantation Procedure
The surgery is performed in a very similar way to a standard pacemaker implantation. Typically, the left side of the chest is selected. Under local anesthesia, a small incision of about 4 to 5 centimeters is made just below the collarbone. One or two wires (leads) are introduced through a vein (the subclavian or cephalic vein). The main defibrillation lead is securely placed in the right lower chamber of the heart. An additional lead may also be placed in the right upper chamber to provide pacing if needed, or to help the device better identify the type of irregular heartbeat.
Once the leads are properly positioned and tested to ensure they function well, they are secured with stitches to the underlying muscle. Next, the leads are connected to the ICD device, which is placed under the skin in a specially created pocket. Finally, the incision is stitched closed, and the patient returns to their room. Patients are usually up and moving either the same day or the following morning, and a single overnight stay at the hospital is typically all that is required. During the first few days, there may be mild pain in the area, which can be managed with standard over-the-counter painkillers.
Potential Complications
The complications associated with an ICD implantation are broadly the same as those for a standard pacemaker.
However, a specific situation that can occur uniquely in ICD patients is a condition called an "electrical storm". In this case, the ICD delivers multiple, consecutive shocks—more than three within a single day—to deal with a severe flare-up of irregular heartbeats. This condition requires immediate hospitalization in a specialized cardiac center, where doctors can take all necessary measures and adjust treatments.
Routine Follow-up and Monitoring
Like standard pacemakers, ICDs are checked externally using specialized portable computers called programmers, which are available at hospitals performing these procedures. Each manufacturer has its own specific programmer. With this device, medical professionals can check how the ICD is working, give commands, and adjust settings as needed. The first check-up usually takes place one month after implantation, and after that, patients are monitored every six months.
During every follow-up appointment, doctors check the condition of the wires, the battery level, and retrieve data from the device’s memory about any irregular heartbeats that occurred and how they were treated. This allows the medical team to understand the patient’s overall heart rhythm health and adjust medications if necessary. Additionally, if the device has intervened to treat an irregular heartbeat, doctors can verify if it functioned correctly.
Adjusting to Life with an ICD
Immediately after the surgery, there may be temporary pain around the incision site, which is easily managed with standard painkillers. Over the days, this initial sensation fades, and most patients completely forget they even have a foreign object in their body. During the first two months, and especially during the first 15 days, patients are advised to restrict aggressive movements of the arm on the side of the device, as the muscle that moves the arm is the one where the wires are anchored.
Another important precaution involves awareness of Electromagnetic Interference (EMI). As mentioned, the ICD monitors the heart’s electrical signals to detect dangerous rhythms and deliver immediate therapy. However, if the device accidentally picks up signals from a strong external electromagnetic source, it can become confused and deliver an unnecessary shock (inappropriate therapy). Standard household electrical appliances do not cause issues. However, direct contact of stray electrical current with the human body can be detected by the device and trigger an inappropriate shock. Fortunately, most modern homes feature safety residual current devices (RCDs/circuit breakers) that automatically cut off power if an electrical leak occurs.
In public spaces, security scanners at banks, department stores, and airports do not cause problems. General maintenance work with power tools is usually safe; however, if it cannot be avoided, using insulating gloves is recommended. Patients with a traditional ICD are strictly prohibited from undergoing an MRI scan. However, in recent years, our center routinely implants MRI-compatible ICDs, effectively removing this potential obstacle.
Device Battery Replacement
Replacing the device is done in a very similar way to standard pacemakers. Typically, ICD batteries (generators) last between 5 and 10 years. Newer devices may offer even longer lifespans.
Subcutaneous ICD (S-ICD)
Traditional ICD systems use transvenous leads, which are wires that travel through the veins and sit inside the heart. Over time, these can lead to infections, meaning the entire system has to be removed. Removing these wires is a difficult and quite risky surgery, as they fuse with the blood vessels and heart tissue over time. To avoid these issues, the Subcutaneous ICD (S-ICD) was developed. This device is placed entirely under the skin, without ever touching the blood vessels or the inside of the heart.
The S-ICD is usually chosen for younger individuals with a long life expectancy. Because they will need multiple battery replacements throughout their life, they face a higher long-term risk of infections or complications from traditional wires. However, the S-ICD cannot provide regular pacing support. Therefore, patients who receive it must not have a medical need for a pacemaker. The surgical procedure usually requires two or three small incisions in the chest area to place the device and the subcutaneous wire along the chest wall.